Abstract

Aseptic loosening, mainly caused by migration, is one of the most common indications for revisions in unicompartmental knee arthroplasty (UKA). In this study, we investigated the early migration of the Persona Partial Knee (PPK, Zimmer Biomet, Warsaw, IN), a cemented medial fixed-bearing unicompartmental knee prosthesis, and evaluated the clinical results. 26 primary PPKs were implanted. Radiographs were obtained direct postoperatively, at 6 weeks, 6, 12 and 24 months postoperatively. Migration of the femoral and tibial component was calculated using model-based radiostereophotogrammetric analysis (mRSA) in terms of translations and rotations. Patient-reported outcome measures (PROMs) were also registered. At 24 months postoperatively, we found low migration of both the femoral and tibial component in the first 6 months, after which both components stabilized. Only the rotation of the tibial component about the z-axis did not stabilize. All PROMs improved after 24 months compared with preoperative PROMs. The Persona Partial Knee shows low migration of both the femoral and tibial component and PROMs were improved at 24 months follow-up. Long-term follow-up is needed to investigate the performance of the prosthesis compared with other prostheses.