Abstract
Modular acetabular systems have several advantages; however, increased polyethylene wear has been recognized as a disadvantage. In our study, an acetabular shell design with a highly polished inner surface, a locking mechanism that minimizes micromotion, and a high polyethylene liner to shell conformity was evaluated. A prospective cohort of 50 consecutive hips was followed up for a mean of 3.6 years (range, 2-7 years). One patient required an early revision for recurrent instability. Mean linear head penetration rates were 0.242 mm/year (range, 0.048-0.655 mm/year). The true linear wear rate was calculated after subtracting the linear wear in the first follow-up year from the total wear at the end of the follow-up. The true linear wear rate was 0.173 mm/year (range, 0.03-0.423 mm/year). A positive correlation was found between wear rate and male gender. The low head penetration rates suggest that a polished liner with an improved locking mechanism and increased liner conformity can reduce polyethylene wear.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.