Abstract

The longevity of valved right ventricle to pulmonary artery (RV-PA) conduits is limited due to calcification and degeneration of non-viable structures. Xenografts are commonly used because of the restricted availability of cryopreserved homografts. Tissue-engineered (de-cellularised) pulmonary valves (TEPVs) were thought to be a valuable alternative to cryopreserved pulmonary homografts due to postoperative seeding with viable autologous vascular endothelial cells. From July 2007 to December 2008, xenogenous TEPV (Matrix P plus) were implanted in 16 patients in the right ventricular outflow tract for different indications, related to congenital heart disease. Mean age at operation was 14+/-11 years, including three patients younger than 1 year. The median conduit size was 22 mm (range: 13-26 mm). The early and late survival rates were 100%. At a median follow-up of 10 months (range: 2-17 months), six patients (38%) had to be re-operated upon due to obstructed grafts. Five of these patients were older than 13 years (range: 13-26 years); one patient was younger than 1 year. On echocardiography before re-operation, mean systolic gradient in the main PA was 66+/-18 mmHg. In explanted conduits, we found a predominantly peripheral conduit narrowing without leaflet calcification. Histological examination revealed stenosis formation, due to inflammatory infiltration and severely fibrogenic pseudo-intimal reaction. On the basis of our short-term results, the Matrix P plus de-cellularised tissue-engineered pulmonary valve cannot be regarded as an ideal conduit for right ventricular outflow tract reconstruction, as the widespread use of these grafts may increase the possibility of frequent early conduit failures.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call