Abstract
Multi-filament orthopaedic cerclage cables have been used in clinical practice for several decades, and are commonly utilized to provide greater trochanteric fixation following an osteotomy or fracture. We present the first known report of patients who experienced early catastrophic failure following use of the Dall-Miles Cable Grip System (DMCGS). A root cause analysis determined that the jaws of one crimper had an increased distance at closure, resulting in inadequate ultimate fixation strength. It was discovered that operating room staff had not been trained to perform regular calibration checks and the required calibration tool was not included in any of the institution’s DMCGS sets. Surgeons should be aware that these surgical instruments require regular maintenance and should be gauge tested prior to every use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.