Abstract

Transcatheter closure of perimembranous ventricular septal defects is challenging. Double disk-designed devices have high adverse event rates; therefore, research into new devices persists. One such devise is the LifetechTM Konar-MF Occluder device (MFO), which is increasingly used. The purpose of this study is to present mid-term results of MFO for transcatheter closure of patients with perimembraneous ventricular septal defect. Records of the 52 patients with perimembraneous ventricular septal defect who had undergone transcatheter closure with MFO were reviewed. Demographic, echocardiographic, and procedure details were investigated. Three years follow-up results were recorded. We closed the perimembraneous ventricular septal defect in 51 of 52 patients (98%). The femoral venous approach was used in 27 patients (53 %) whilst no arteriovenous loop was established in the remaining patients. No significant procedure-related complication occurred. On day 1 echocardiography, the residual ventricular septal defect rate was 31%. Mean±SD (range) follow-up duration was 36 ± 7,9 (18-54) months. In the follow-up, at 6th months, only four patients had hemodynamically insignificant residual defects. No severe dysrhythmia was detected including complete heart block. A right bundle branch block pattern was seen in one patient. This study showed that MFO is a safe and effective device in the transcatheter treatment of perimembraneous ventricular septal defect's with mid-term follow-up.