Abstract
Pulse oximetry normally provides a reliable indicator of arterial blood oxygen saturation. Individuals with a variant hemoglobin with low oxygen affinity would exhibit pulse oximetry values consistent with hypoxemia despite maintaining adequate oxygen delivery to tissues. We describe a patient we diagnosed with hemoglobin Rothschild and his mother. He presented with SpO2 81% on pulse oximetry and PaO2 on arterial blood gas of 47 mm Hg, while his mother had SpO2 87%. These values were significantly lower than physical examination suggested. A possible dyshemoglobinemia should be suspected when discordance exists between pulse oximetry readings and physical examination findings.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.