Abstract

Conventional digital tomosynthesis (DTS) reconstruction by using the filtered-backprojection algorithm requires a full field-of-view scan and relatively dense projections to obtain high-quality images, which results in a high radiation dose to patients. Interior DTS (iDTS) with a proper collimator offers a possible imaging modality for reducing the dose of radiation delivered because the X-ray beam is able to target a small region-of-interest (ROI) containing the target area. Collimators for iDTS often have a fixed rectangular shape, but focusing the X-ray beam on an arbitrarily shaped ROI would be preferable because it further reduces the excessive radiation dose. In this study, we propose a new iDTS scan method to create an ROI with an arbitrary shape to minimize the radiation dose at each angle of view. We used a compressed-sensing-based algorithm for accurate iDTS reconstruction. To validate the proposed method, we performed a systematic simulation and an experiment, and we investigated the image characteristics. Our results indicate that the proposed method may effectively reduce radiation dose in iDTS in real imaging systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.