Abstract

The objective evaluation of thermal and humidification processes in the pulmonary system requires accurate dynamic measurements of temperature and water vapor concentration of a flowing gas mixture. The adequacy of instruments used for such measurements can only be determined by dynamic calibration techniques. We have developed a method of producing step changes in temperature and water vapor content of a gas mixture undergoing controlled steady flow. The system consists of two reservoirs and a slide valve that switches a test section between them. The inlet (usually a probe or catheter tip) of the device to be calibrated is positioned in the test section. The flow rate through the test section is minimally changed during the transition between gas from one reservoir to that of the other. The system has been used to analyze the response of a thermistor and a respiratory mass spectrometer to changes in gas temperature and water vapor.