Abstract

Methods Our MR-compatible phantom resembles the anatomy of the heart and of the thoracic vessels of a 60 kg subject. Water flow (2-4 l/min) is driven into the system by a pump located outside the MR room. Gadolinium injections (Gadovist, gadobutrol, Bayer Schering, Germany) are performed in the vena cava by a power-injector (Medrad Solaris, Germany). Progressive dilution of the Gadolinium bolus across the chambers generates the AIF. A fraction of the cardiac output perfuses two myocardial compartments (made by arrays of parallel tubes) where perfusion can be controlled precisely and independently. Outside the MR room, a control unit allows for precise measurement and control of cardiac output and myocardial perfusion. The phantom was tested in a 3T-scanner (Philips Achieva, Netherlands). Acquisition of the perfusion sequence was repeated multiple times with a k-t SENSE sequence (flip-angle=20°; TR2.2 ms; TE1.1 ms; k-t factor 5, 11 training profiles) using similar flow conditions to demonstrate the reproducibility of the measurements. Different flow conditions were used to assess the response of the system to different myocardial perfusion rates. Quantification was performed by Fermi deconvolution.

Highlights

  • Development of quantitative MR-perfusion would benefit from the availability of a MR-compatible harware phantom allowing for a realistic simulation of the AIF and of the myocardial signal intensity curves

  • Our MR-compatible phantom resembles the anatomy of the heart and of the thoracic vessels of a 60 kg subject

  • Water flow (2-4 l/min) is driven into the system by a pump located outside the MR room

Read more

Summary

Objective

To describe a novel hardware MR-perfusion phantom to model first-pass of a bolus of contrast in the cardiac cavities and large thoracic vessels (arterial input function, AIF) and in the myocardium, allowing for the acquisition of dynamic first-pass MR-perfusion data and for a precise control of cardiac output and myocardial perfusion for true validation of perfusion quantification

Background
Methods
Results
Conclusions

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.