Abstract

Prototypes of radiation detectors for clinical dosimetry were constructed using commercially available synthetic diamond. Polished diamond plates of detector grade and standard quality were used. The metallization scheme was a typical Ag–Ag contact for the detector grade diamond and a new metallization technique for the standard quality diamond. This was for determining which detector is more appropriate for radiotherapy dosimetry needs. The detector signal was analyzed as a function of bias voltage in the 0–500 V range both in absence of radiation and in a Co-60 gamma-ray beam. The response linearity was assessed in Co-60 gamma-ray beam for dose-rates in the range from 0.2 to 1.2 Gy min −1. Pre-irradiation effects were also investigated. Signal rise time, signal stability and reproducibility are reported and discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.