Abstract
A system for measuring parameters of ventilation and respiratory mechanics in intact anesthetized hamsters was developed. Means +/- SD of five weekly measurements in eight hamsters were 0.68 +/- 0.09 ml for tidal volume, 45 +/- 14 breaths.min-1 for respiratory frequency, 29.3 +/- 10.4 ml.min-1 for minute ventilation, 0.301 +/- 0.080 ml.cmH2O-1 for dynamic compliance, 0.435 +/- 0.151 cmH2O.ml-1.s for inspiratory resistance, 0.311 +/- 0.101 cmH2O.ml-1.s for expiratory resistance, and 0.334 +/- 0.096 cmH2O.ml-1.s for average resistance. The standard deviations of five consecutive measurements of these parameters on the same day were typically 10% of the corresponding mean values suggesting that acute changes as small as 10% can be detected in a study of similar size. The corresponding value for five single measurements on different days at weekly intervals was 30%. We believe that this system yields reasonable and repeatable measurements, and, because it is atraumatic, it can be used in chronic studies of diseases in hamsters and other small animals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of applied physiology: respiratory, environmental and exercise physiology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.