Abstract

A radial projection sliding-window sequence has been developed for imaging the rapid flow of (3)He gas in human lungs. The short echo time (TE) of the radial sequence lends itself to fast repetition times, and thus allows a rapid update in the image when it is reconstructed with a sliding window. Oversampling in the radial direction combined with angular undersampling can further reduce the time needed to acquire a complete image data set, without significantly compromising spatial resolution. Controlled flow phantom experiments using hyperpolarized (3)He gas exemplify the temporal resolution of the method. In vivo studies on three healthy volunteers, one patient with chronic obstructive pulmonary disease (COPD), and one patient with hemiparalysis of the right diaphragm demonstrate that it is possible to accurately resolve the passage of gas down the trachea and bronchi and into the peripheral lung.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.