Abstract
A system consisting of a high-pressure tolerant microfluidic glass chip, high-speed absorbance imaging, and image processing has been developed to study rapid dynamic events like pH change in a multiphase flow. The system gives both kinetic and quantitative equilibrated information. By tracking the interactions of aqueous additive mixtures and liquid CO2, at 80 bar and 24 °C, under flow, measurement at a given P, T condition is done in 0.25 s. The acidification rate to steady state was found to be mass transport limited, occurring in less than 1 s. For 30 mM of the additives ammonium acetate and ammonium formate, equilibrium pH of 4.5 and 4.1, respectively, was seen. These additives are of key importance in common mobile phases used in SFC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.