Abstract

We describe a new dynamic optical tomography system that is, unlike currently available analog instrumentation, based on digital data-acquisition and filtering techniques. At the heart of this continuous wave instrument is a digital signal processor (DSP) that collects, collates, processes, and filters the digitized data set. A digital lock-in filter that has been designed for this particular application maximizes measurement fidelity. The synchronously-timed processes are controlled by a complex programmable logic device (CPLD) that is also used in conjunction with the DSP to orchestrate data flow. Real-time data rates as high as 140Hz can be achieved. The operation of the system is implemented through a graphical user interface designed with LabVIEW software, Performance analysis shows very low system noise (~600fW RMS noise equivalent power), excellent signal precision (<0.04% - 0.2%) and long-term system stability (<1% over 40 min). A large dynamic range (~195dB) accommodates a wide scope of measurement geometries and tissue types. First experiments on tissue phantoms show that dynamic behavior is accurately captured and spatial location can be correctly tracked using this system.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call