Abstract
Flexometallic tubes are kink-resistant and extensively used in neurosurgeries. The manufacturers of flexometallic tubes recommend single usage. In low-income countries, due to the cost restraints, flexometallic tubes are commonly reused after ethylene oxide sterilization. With this technique, flexometallic tubes are effectively sterilized, but the physical integrity of the endotracheal tube cuffs may be compromised.1 Tube kinking2 and intraluminal blebs due to dissection3 are observed with the reused flexometallic tubes. These obstructions lead to difficulty in ventilation and inability to pass a suction catheter freely. Flexometallic tubes should be thoroughly checked for external deformity, cuff integrity, and tube patency for safe reuse.Despite thorough assessment of reused flexometallic tubes, the partial unfurling of nylon spirals with distal herniation into the tube (fig.) may still happen intraoperatively, leading to dynamic obstruction resulting in variable resistance during suction catheter negotiation (see Supplemental Digital Content 1, video S1, http://links.lww.com/ALN/C848). Along with the heavy weight of the circuit, the varied positions of head and neck result in additional rotational bending of flexometallic tubes, contributing to partial unfurling, which may not cause complete airway obstruction but can lead to difficult ventilation requiring airway change. The simple technique of passing a suction catheter through flexometallic tubes can help in identifying such partial tube obstruction. Reinforced tracheostomy tubes are also frequently reused after a thorough cleaning and visual inspection.Adhering to the manufacturer’s recommendation regarding reuse is the best way to prevent these complications. However, when clinically necessary and under the right circumstances only, flexometallic tubes should be reused after a thorough assessment.The authors are thankful to the patient who consented to publish the current article. The authors are also thankful to all anesthesia technicians who helped them ensure patient safety.The authors declare no competing interests.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.