Abstract

In the optical imaging system, the dynamic modulation transfer function (DMTF) is affected by the wave aberration of optical system, the signal resolution and transfer characteristics of detector modules, and the image blur caused by the carrier motion, which has become one of the important parameters of the photoelectric imaging system. Based on the testing principle of dynamic and static modulation transfer function of photoelectric imaging system, a DMTF measuring device was developed, in which the focal length of optical collimation system was 10 000 mm, and the speed control range of moving target was 30 mm/s~5 000 mm/s, which met the DMTF test of photoelectric imaging system with long focal length. A series of experiments were carried out, and the results show that the repeatability of DMTF measurement is better than 0.01, and the uncertainty of measurement can reach to U= 0.05 (k=2).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.