Abstract

Traditional we use the standard accelerometer to calibrate the dynamic force. The standard accelerometer was calibrated by vibration standard machine. However, the calibration condition is not as the use condition. So we need to develop a dynamic force calibration method, which was the same condition as used. This paper describes a new method of dynamic force calibration. Now we measure the acceleration by a laser Interferometer. Traceability for force is realized via the measurement of acceleration with laser-Doppler-interferometers and the determination of the dropping masses. This paper introduces the structure and the working principle of the equipment, performance parameters and the uncertainty evaluation. This paper also compares the data between the accelerometer and laser-Doppler-interferometer.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.