Abstract

Cryogen spray cooling of the tissue surface was investigated for laser welding applications. Benefits include reduced thermal damage to the papillary dermis and reduced operation time. Two-cm-long, full-thickness incisions were made on the backs of guinea pigs, in vivo. India ink was used as an absorber and clamps were used to appose the incision edges. Continuous-wave, 1.06-μm, Nd:YAG laser radiation was scanned over the incisions, producing ~100 ms pulses. A 4-mm-diameter laser spot was used with a constant power of 16 W. The total operation time was 60 or 120 s. Cryogen was delivered in spurt durations of 20, 60, or 100 ms, with 2 or 4 s between spurts. The working distance was approximately 12 cm, and the spray covered an area of about 5.0 x 5.0 cm. Control welds were irradiated for 20, 40, or 60 s. Total operation times were reduced from 10 min without dynamic cooling to 1 min with dynamic cooling. Optimal tensile strength was 1.7 ± 0.7 kg/cm 2 , comparible to stengths of 2.1 ± 0.7 kg/cm 2 reported in previous studies without cryogen cooling (p>0.25). Thermal damage in the papillary dermis measured 320 ± 80 μm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.