Abstract

Purpose: The accuracy of Goldmann applanation tonometry (GAT) has been shown to depend on several biomechanical properties of the cornea. Newer tonometry devices (e.g., the Dynamic Contour Tonometer PASCAL? [DCT] and the Tono-Pen? XL [TP]) have been designed to accurately measure intraocular pressure (IOP) independent of corneal thickness (CCT) and pathology. This study investigates the influence of corneal pathologies on the accuracy of these IOP measuring devices, and compares this accuracy to that of direct intracameral IOP measurement. Methods: 8 eyes of 8 patients suffering from corneal pathologies scheduled for penetrating keratoplasty, and 10 eyes of 10 patients scheduled for cataract surgery (control group) were examined. Before surgery, the anterior chamber was cannulated at the temporal corneal limbus. In a closed system, the intraocular pressure (IOP) was directly set to 10, 20, and 30 mmHg with a manometric water column. Intraocular pressure measurements taken by GAT, DCT, and TP were compared to intracameral measurements obtained by a precision reference pressure sensor. Results: Control group: All three methods showed good agreement with the intracameral readings (mean deviation of all three devices, -0.9 mmHg). Group with corneal pathologies: The TP yielded the most exact IOP values in the group with corneal pathologies when taking all diagnoses into account. The mean deviations from the intracameral IOP measurements were -0.9 mmHg ± 3.2 mmHg (mean ± SD) for TP, -2.9 mmHg ± 3.3 mmHg for GAT, and -5.0 mmHg ± 7.9 mmHg for DCT. For bullous keratopathy, the most exact IOP readings were obtained by the TP (mean deviation -0.2 mmHg ± 3.5 mmHg). The TP and GAT devices underestimated IOP in the patients with Fuchs’ endothelial dystrophy; all 3 devices underestimated adjusted IOP after keratoplasty. DCT showed the greatest deviations from adjusted IOP in the case of non-herpetic scars. In the control group, none of the devices showed a statistically relevant dependency on CCT. Nevertheless, in the group with corneal pathologies, only TP showed no dependency on CCT. Conclusion: Our results suggest that the Tono-Pen XL? is the most accurate measurement device to determine IOP in patients with corneal pathologies, especially in patients suffering from corneal edema (bullous keratopathy). GAT yielded surprisingly exact IOP values in patients suffering from irregular corneal surface. DCT showed a high degree of deviation from the adjusted IOP, and should not be used to determine IOP in corneas with the disorders listed here.

Highlights

  • How to cite this paper: Neuburger, M., et al (2014) Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral Intraocular Pressure (IOP) Measurements in Patients with Corneal Pathologies

  • Our results suggest that the Tono-Pen XL® is the most accurate measurement device to determine IOP in patients with corneal pathologies, especially in patients suffering from corneal edema

  • Group with corneal pathologies (Figure 3): When taking into account all examined patients, the Tono-Pen® XL (TP) yielded the most exact IOP values in the group with corneal pathologies; we found a mean deviation of −0.9 mmHg ± 3.2 mmHg

Read more

Summary

Introduction

How to cite this paper: Neuburger, M., et al (2014) Dynamic Contour Tonometry, Tono-Pen XL®, and Goldmann Applanation Tonometry in Comparison to Intracameral Intraocular Pressure (IOP) Measurements in Patients with Corneal Pathologies. DCT showed the greatest deviations from adjusted IOP in the case of non-herpetic scars. None of the devices showed a statistically relevant dependency on CCT. In the group with corneal pathologies, only TP showed no dependency on CCT. Conclusion: Our results suggest that the Tono-Pen XL® is the most accurate measurement device to determine IOP in patients with corneal pathologies, especially in patients suffering from corneal edema (bullous keratopathy). DCT showed a high degree of deviation from the adjusted IOP, and should not be used to determine IOP in corneas with the disorders listed here

Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.