The differences in dynamic thermal changes during laser lithotripsy between various laser pulse modes are unclear. We used thermography to evaluate the temporal changes in high-temperature areas during laser activation in order to compare different laser pulse modes. An unroofed artificial kidney model was used for the experiments. The laser fired for 60 s with a laser setting of 0.4 J/60 Hz in the following four different laser pulse modes without saline irrigation: short pulse mode (SPM), long pulse mode (LPM), virtual basket mode (VBM) and Moses mode (MM). Using the first 30 s of moving images, we compared the ratio of a high-temperature area of >43 °C to the total area every 5 seconds. The dynamic changes in fluid temperatures were shown to be different between the laser pulse modes. The extent of the high-temperature areas during the laser activation was large in the LPM and MM compared with the SPM and VBM. While the high-temperature areas expanded in an anterior direction in the early laser irradiation period using the LPM, they spread in a posterior direction in the early laser activation period using the MM. Although only the temperature profile in one specific plane was investigated, these results are considered useful for preventing thermal injuries during retrograde intrarenal surgeries.