Abstract

We used a novel 3-dimensional (3D) echocardiographic technique to evaluate the impact of a coronary sinus-based percutaneous transvenous mitral annuloplasty (PTMA) on dynamic changes in mitral annular geometry and motion during the cardiac cycle in 8 sheep with ischemic mitral regurgitation. Using real-time 3D echocardiographic data before and after PTMA, 10 points along the saddle-shaped annulus were identified. For every 3D volume/frame during a cardiac cycle, we assessed mitral annular area and excursion defined as the traveling distance of the annular center. The PTMA device reduced both minimum and maximal mitral annular area (9.5 +/- 0.9-7.0 +/- 0.6 and 12.8 +/- 1.3-9.8 +/- 1.5 cm(2), P < .001 for both, respectively) with reduction of mitral regurgitation jet area (5.1 +/- 2.3-1.2 +/- 0.8 cm(2), P < .001), whereas it did not significantly impair mitral annular excursion amplitude (8.3 +/- 1.1-7.0 +/- 1.9 mm, P = .13). This 3D echocardiographic method noninvasively enabled dynamic study of mitral annular geometry and motion with quantitative analysis of the impact of PTMA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.