Abstract This paper proposes a new dynamic calibration procedure of the force measuring system used in instrumented impact test apparatus. The method involves calculation of the force calibration factor on the basis of energy values obtained executing the Charpy or Izod test for a pendulum device and on the basis of velocity measurements in a rebound test for the falling dart impact apparatus. The procedure is simple, quick, and has the advantage that calibration is performed under the same dynamic conditions as in a test itself. Moreover, it does not require any additional equipment other than that used for conventional impact test execution: an impact device, an instrumented tup, and a data acquisition system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.