Abstract

Dual energy X-ray absorptiometry (DXA) was used to measure human body composition on a high speed fan beam scanner. High and low energy attenuation pairs, produced by the various combinations of fat mass and fat-free mass in the human body, were compared to attenuation values produced by standard materials (aluminum and acrylic). These standards were measured in various combinations to construct calibration curves for fat and fat-free mass. Primary calibration of the aluminum/acrylic combinations was achieved by direct comparison to the dual energy attenuation produced by stearic acid and pure water. Whole body examinations were accomplished using three 45 s longitudinal passes of the fan beam. These passes were acquired and assembled to create a giant, isocentric fan beam with a single center of focus. In vivo precision was 0.009 g/cm 2 for BMD and 425 g for fat mass and fat-free mass (SD).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.