Abstract

In current electrodiagnostic criteria for chronic inflammatory demyelinating polyneuropathy, the cutoff values of distal compound muscle action potential (DCMAP) duration are defined using electromyogram low-cut filter setting of 20 Hz. We aimed to assess effects of low-cut filter on DCMAP duration (10 vs. 20 Hz). We prospectively measured DCMAP duration in 130 normal controls and 42 patients, fulfilling diagnostic criteria for typical chronic inflammatory demyelinating polyneuropathy by European Federation of Neurological Societies/Peripheral Nerve Society. Distal compound muscle action potential duration was significantly shortened with 20-Hz than 10-Hz filtering. When the cutoff values were defined as the upper limit of normal (ULN, mean + 2.5SD), the sensitivity/specificity was 67%/95% in 10-Hz recordings, and 69%/95% in 20-Hz recordings. This diagnostic accuracy was similar to that defined by receiver operating characteristic analyses. Distal compound muscle action potential duration significantly affected by the low-cut electromyogram filter setting, but with at least 10 and 20 Hz, the diagnostic accuracy is similar.