Abstract

AbstractIn relation to sterilization of medical supplies, the degree of degradation by γ‐ray and electron beam irradiations of homopolypropylene (HP), copolypropylene (CP, coplymer including 6% of ethylene unit) and polymethylpentene were compared, and chemiluminescence (CL) of irradiated polymers were measured. HP degraded extremely around the sterilization dose (2.5 Mrad) by either γ‐ray or electron beam irradiations. In the case of CP and polymethylpentene, stabilities of polymers far differed between γ‐ray and electron beam irradiations. The polymethylpentene was more stable than the polypropylenes against irradiation. The counts of CL emitted by recombination of peroxy radical (ROO ·) increased with increasing dose, reflecting degrees of oxidation of polymers. The degradation of polymers was independent of irradiation sources, rather it depended on the degree of oxidation. It was found that CL analysis are favorable for estimation of degradation in irradiated polymers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.