Abstract

Several thermal treatments were studied to improve the TL dosimetric characteristics of the newly developed and highly sensitive LiF:Mg,Cu,Si sintered discs. A treatment of 260 ∘ C for 10 min was found to be able to fully restore the TL sensitivity and glow curve structure of the dosimeters which were otherwise altered by heating to 300 ∘ C . A dual-step thermal treatment of 300 ∘ C for 10 min followed by 260 ∘ C for 10 min was found to be optimum and it also minimized the unwanted glow peaks at higher temperatures, thereby reducing the residual TL signal to about 0.025%. For the reuse, annealing treatment of 260 ∘ C for 10 min was found to cause no noticeable change in the TL sensitivity and glow curve structure for 30 cycles of anneal and readout. An effort was also made to understand the role of dual-step thermal treatment in the light of known conversion of Cu + and Cu 2 + and the changes in precipitates and defect complexes involving Mg, Cu, Si, O and OH ions and their dissolution in LiF by heating.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.