Abstract
This paper describes a method for single exposure contrast-enhanced dual-energy imaging of tumors utilizing a scanned multi-slit system for digital mammography. This photon counting system employs an array of silicon strip detectors in an edge-on geometry. In the multi-slit setup, the line detectors and pre-collimator slits are aligned orthogonal to the scan direction. This geometry is advantageous to dual-energy imaging, since it allows differential filtering of the x-ray beam in the pre-collimator slits. A high-energy image is constructed from those lines where the filter material has been chosen to harden the x-ray beam and the low-energy image from the lines with a filter producing softer beams. Both images are obtained in the same scan, eliminating the need to change tube voltages and anode materials and minimizing the risk of motion artifacts. The method is illustrated on a purpose-built phantom and logarithmic subtraction of the images produces images essentially free of anatomical clutter with the contrast-enhanced targets clearly visible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.