Abstract

Remote Raman instruments have become powerful analytical tools in some special environments. However, ambient daylight is the main limitation in the data acquisition process. To suppress daylight background interference and obtain a high signal-to-background ratio (SBR), we develop a dual-wavelength wide area illumination Raman difference spectroscopy (WAIRDS) system for daytime remote detection. In the WAIRDS system, a wide area illumination scheme and shifted-excitation Raman difference spectroscopy method are used to improve the reliability of collected Raman spectra. Measurements of polystyrene indicate that the WAIRDS system can be operated to obtain background-free Raman spectra under different levels of daylight background interference. The remote results show that the improvement in SBR is about three- to fivefold, and the system can work at distances of up to 9.2 m on a sunny afternoon. Moreover, to be close to the actual detection, the system is used for mineral and explosive raw material detection during daytime measurement. Measurements show that the WAIRDS system will be a useful tool for many remote applications in the future.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.