Abstract
The absorption spectra of effluent dialysate in the ultraviolet region were investigated. A novel dual-wavelength spectrophotometric method for uric acid determination in effluent dialysate and an optoelectronic sensor based on UV LED were developed. Clinical trials of the proposed sensor were carried out in the dialysis unit of St. Petersburg Mariinsky Hospital. The relative error of measurement for the concentration of uric acid does not exceed 10%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.