Abstract

The preferred use of dual-coil implantable defibrillator lead systems in current implantable defibrillator therapy is likely based on data showing statistically lower defibrillation thresholds with dual-coil defibrillator lead systems. The following review will summarize the clinical data for dual- versus single-coil defibrillator leads in the left and right pectoral implant locations, and will then discuss the clinical implications of single- versus dual-coil usage for atrial defibrillation, venous complications, and the risks associated with lead extraction. It will be noted that there are no comparative clinical studies on the use and outcomes of single- versus dual-coil lead systems in implantable defibrillator therapy over a long-term follow-up. The limited long-term reliability of defibrillator leads is a major concern in implantable defibrillator and cardiac resynchronization therapy. A simpler single-coil defibrillator lead system may improve the long-term performance of implanted leads. Furthermore, the superior vena cava coil is suspected to increase interventional risk in transvenous lead extraction. Therefore, the need for objective data on extractions and complications will be emphasized.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call