Abstract

We have developed a programmable Josephson voltage standard that can produce voltages up to 20 V with a resolution of better than 0.1 µV over the whole voltage range and better than 1 nV for voltages up to 10 mV. The standard has two superconductor–normal metal–superconductor junction arrays connected in series and driven by two radiofrequency oscillators. The cryogenic part of the standard is based on a cryocooler. The new standard agrees with the primary quantum voltage standard maintained at the National Measurement Institute, Australia, within 10 nV and forms the basis of an automated calibration system for digital multimeters and voltage references.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.