Abstract
A dual simultaneous photometric-conductometric detector for microfluidic chip is reported. Two different detect methods confocal LIF and moveable C 4D were combined together. They shared a common detection cell and could respond simultaneously. The dual detector offered possibility of measurement one analyte in two different ways and was advantageous in analyses of mixtures containing organic and inorganic ions. To improve the separation efficiency, pinch injection was carried out by a series of electrokinetic manipulations. Rhodamine B was used to evaluate the performance of the dual detector. The dual detector had highly sensitivity and could offer simultaneous information. © 2009 SPIE.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.