Abstract

To biomechanically compare the stiffness of midshaft synthetic clavicle osteotomies fixed with either superior anatomic pre-contoured locking plates, anterior anatomic pre-contoured locking plates, or short-segment dual orthogonal mini-plate fixation. Controlled laboratory study. Specimens Twenty-one synthetic pre-osteotomized clavicles were separated into three groups: superior plating, anterior plating, or dual-plating. Each clavicle was sequentially tested in non-destructive cycles of axial compression, three-point bending, and torsion. Load and displacement were recorded. Stiffness was calculated. No statistically significant differences were found between construct stiffness during axial compression, three-point bending, or torsional testing. One superior plated clavicle suffered catastrophic failure during axial compression. One dual mini-fragment plated clavicle suffered catastrophic failure during torsion. Orthogonal dual mini-fragment fixation of transverse clavicle fractures is biomechanically similar to superior and anterior pre-contoured anatomic locking plate fixation. No statistically significant differences in construct stiffness were found in axial compression, three-point bending, or torsion testing. Further clinical research is required to determine the long-term stability of dual mini-fragment plate fixation. IV.