Abstract

We present a fluidic micro-calorimeter that shows high versatility measuring both thermal properties of samples and enthalpies of bio-chemical processes. The micro-calorimeter consists of two stacked chips with silicon nitride membranes that define the liquid chamber of 2.6 μl. The chamber has two inlets which allow real time reactions monitoring. We measured the thermal conductivity and diffusivity of water-methanol solutions with 1% resolution, and we determined the mixing enthalpy for watermethanol system with a signal/noise ratio of 20.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.