Abstract

To describe the experience of coronary chronic total occlusions (CTOs) percutaneous coronary interventions (PCI) using antegrade fenestration and re-entry (AFR) technique with a dedicated dual guidewire balloon (DGB). Antegrade dissection and re-entry (ADR) techniques has been emphasized in recent worldwide CTO consensus documents. We investigated the feasibility and safety of DGB as a dedicated device to perform guidewire-based AFR. Fourteen consecutive patients with complex CTO (J-CTO score: 3.1 ± 0.9) underwent DGB-AFR in the years 2020-2021. DGB-AFR consists in advancing the DGB over a guidewire that reached the vessel distal to the CTO in an extra plaque fashion, inflating/deflating the DGB to create fenestration between subintimal space and the true lumen and advancing a proximal re-entry guidewire through fenestration in the true lumen. DGB-AFR alone was successful in 10 of 14 (71%) cases, a rescue wire-based ADR was needed in two cases for re-entry into the true lumen with a total success rate in 12 of 14 (86%) cases. Among all DGB-AFR cases, four (28%) were performed as a first-line strategy while the remaining 10 (71%) cases were performed as a bail-out strategy after failure of other antegrade crossings for 30 min of procedural time. No DGB-related complications were observed. DGB-AFR is a user-friendly reliable strategy for the treatment of many CTO lesions. It can be used as bail-out after failure of conventional antegrade wiring techniques, achieving high procedural success rate and low occurrence of procedural adverse events.