A data acquisition system has been developed to collect camera events simultaneously with continually digitized electrocardiograph signals and respiratory flow measurements. Software processing of the list mode data creates more precisely gated cardiac frames. Additionally, motion blur due to heart movement during breathing is reduced by selecting events within a specific respiratory phase. Thallium myocardium images of a healthy volunteer show increased definition. This technique of combined cardiac and respiratory gating has the potential of improving the detectability of small lesions, and the characterization of cardiac wall motion.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.