Abstract

Quantitative ultrasound (US) measurements have been suggested for screening of osteoporosis. However, soft tissues overlying bones affect reliability of the measurements. In this in vitro study, a novel dual frequency ultrasound (DFUS) technique is introduced for elimination of the errors induced by soft tissues on bone pulse‐echo US measurements. In DFUS, US reflection from soft tissue‐bone interface is measured with two different US frequencies. By knowing the frequency specific US attenuation and speed in adipose and lean tissues, the effect of soft tissue can be determined. DFUS, conducted at frequencies of 2.25 MHz and 5.0 MHz, was validated using human trabecular bone samples (n = 25) covered with heterogeneous soft tissues. DFUS, reduced (p < 0.01) the mean error induced by soft tissues from 58.6% to ‐4.9% and from 127.4% to 23.8% in broadband ultrasound backscattering and integrated reflection coefficients (at 5.0 MHz), respectively. Our results suggest that DFUS is a technique capable to minimize...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.