Abstract

1. There are few data regarding the accuracy of Hologic QDR-1000W dual-energy X-ray absorptiometry for the measurement of body composition. In two studies, one in an in vitro experimental system using oil and water mixtures and the other in samples of pork meat, the effect of depth and tissue thickness on the measured composition was assessed. In the latter study the measured fat mass was compared with that measured by direct analysis. 2. All data indicated a trend in the measured fat mass with depth, such that more fat was measured at extremes of depth (< 10 cm and > 25 cm) than at intermediate depths. 3. In samples of meat weighing approximately 55 kg, dual X-ray absorptiometry significantly under-estimated the absolute fat mass compared with direct analysis (mean 20.4 +/- 1.65%) by 5-8% or 1-4 kg of fat. 4. These findings are of direct relevance to both clinical and research work using this technique to measure body composition, in particular in circumstances in which changes in body composition and/or tissue thickness are anticipated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.