Abstract

Disinfectants for medical devices are uncommonly a cause of iatrogenic adverse effects. Nevertheless, when misused, they can induce severe complications. Three cases of acute respiratory distress in newborns probably induced by glutaraldehyde are reported. Three children born by Caesarean section between 8 and 19 May 1999 in the same hospital presented acute respiratory distress requiring hospitalization in the neonatal intensive care unit; one child was premature. The clinical appearance, which was initially normal, deteriorated with a respiratory distress in 30 to 60 minutes. Recovery was uneventful in all cases. The diagnosis considered was a hyaline membrane disease. The enquiry conducted after this cluster onset identified, as a main contributing factor, the disinfection procedure recently introduced in the surgical theater. Review of toxicologic data on glutaraldehyde shows this is a highly irritating chemical for the respiratory tract, even at low concentrations. Clinical and radiologic features in these three neonates are compatible with a pulmonary sub-edema on an immature alveolar setting. The hypothesis proposed is that glutaraldehyde, the active ingredient of the biocidal formula used to disinfect the respiratory masks, was massively desorbed from the rubber and foam of which masks are made.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.