Abstract
Here we demonstrate that the film refractive index (RI) can be an even more important parameter than film thickness for identifying nonfouling polymer films to undiluted human blood plasma and serum. The film thickness and RI are two parameters obtained from ellipsometry. Previously, film thickness has been correlated to ultra-low fouling properties. Practically, the film RI can be used to characterize polymer density but is often overlooked. By varying the water content in the surface-initiated atom transfer radical polymerization of zwitterionic carboxybetaine, a minimum of ∼1.5 RI units was necessary to achieve <5 ng/cm(2) of adsorption from undiluted human serum. A model of the film structure versus water content was also developed. These results point to an important parameter and simple approach for identifying surface coatings suitable for real-world applications involving complex media. Therefore, ultra-low fouling using a thin film is possible if it is densely packed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.