To evaluate MR safety at 3 Tesla for a drug eluting coronary stent. A drug eluting coronary stent (Endeavor, cobalt alloy, Medtronic Vascular, Santa Rosa, CA) was evaluated for magnetic field interactions, heating, and artifacts at 3 Tesla. MRI-related heating was assessed with the stent in a gelled saline-filled phantom using a transmit/received RF body coil with a whole body averaged SAR of 2.0 W/kg. Artifacts were characterized using T1-weighted, spin echo, and gradient echo pulse sequences. The stent exhibited minor magnetic field interactions that will not cause migration. Heating was not substantial (+0.5 degrees C). Artifacts may create a problem if the area of interest is in the same area or close to the stent (e.g., for a T1-weighted, spin echo pulse sequence, within approximately 16 mm; for a gradient echo pulse sequence, within approximately 23 mm). Conclusion. The findings indicated that it would be safe for a patient with this cobalt alloy-based, drug-eluting coronary stent to undergo MRI at 3 Tesla or less. Importantly, because of the relative lack of magnetic field interactions, MRI may be performed immediately after implantation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.