Abstract
To evaluate MR safety at 3 Tesla for a drug eluting coronary stent. A drug eluting coronary stent (Endeavor, cobalt alloy, Medtronic Vascular, Santa Rosa, CA) was evaluated for magnetic field interactions, heating, and artifacts at 3 Tesla. MRI-related heating was assessed with the stent in a gelled saline-filled phantom using a transmit/received RF body coil with a whole body averaged SAR of 2.0 W/kg. Artifacts were characterized using T1-weighted, spin echo, and gradient echo pulse sequences. The stent exhibited minor magnetic field interactions that will not cause migration. Heating was not substantial (+0.5 degrees C). Artifacts may create a problem if the area of interest is in the same area or close to the stent (e.g., for a T1-weighted, spin echo pulse sequence, within approximately 16 mm; for a gradient echo pulse sequence, within approximately 23 mm). Conclusion. The findings indicated that it would be safe for a patient with this cobalt alloy-based, drug-eluting coronary stent to undergo MRI at 3 Tesla or less. Importantly, because of the relative lack of magnetic field interactions, MRI may be performed immediately after implantation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.