Abstract

A micro-flow chemiluminescence (CL) system in vivo for glucose determination by the on-line microdialysis sampling is described in this paper. The micro-flow CL system uses discrete sample droplets, which formed at the tip of the capillary with the sampling volume of 4.5 microl. The sol-gel method is introduced to co-immobilize horseradish peroxidase (HRP) and glucose oxidase (GOD) on the inside surface of the micro-flow cell which was fabricated in polymethyl methacrylate (PMMA). The CL detection involved enzymatic oxidation of glucose to D-gluconic acid and H2O2, then H2O2 oxidizing luminol to produce CL in presence of HRP. The microdialysis probe was utilized for sampling in the rabbit blood; the sample throughput was 20 h(-1). The glucose level in blood of the rabbit was on-line monitored with good results.