Abstract

AbstractNew data are presented on drop size distribution at high dispersed phase fractions of organic‐in‐water mixtures, obtained with a light back scattering technique (3 Dimensional Optical Reflectance Measurement technique, 3D ORM). The 3D ORM technique, which provides fast, in‐situ and on‐line drop distribution measurements even at high concentrations of the dispersed phase, is validated using an endoscope attached to a high‐speed video recorder. The two techniques compared favourably when used in a dispersion of oil (density (ρ) = 828 kg m−3, viscosity (µ) = 5.5 mPa s, interfacial tension (σi) = 44.7 mN m−1) in water for a range of 5–10% dispersed phase fractions. Data obtained with the ORM instrument for dispersed phase fractions up to 60% and impeller speeds 350–550 rpm showed a decrease in the maximum and the Sauter mean drop diameters with increasing impeller speed. Phase fractions did not seem to significantly affect drop size. Both techniques showed that drop size distributions could be fitted by the log‐normal distribution. Copyright © 2005 Society of Chemical Industry

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.