Abstract

The hematocrit effect is a hurdle for successful introduction of the dried blood spot (DBS) in a regulated environment. Recently, attempts were taken to overcome the hematocrit effect by whole-cut DBS analysis. This paper presents the next-generation whole-cut DBS; dried matrix on paper disks (DMPD). DMPD eliminated the hematocrit effect and demonstrated better accuracy and precision than regular DBS with partial punching. Observed accuracy and precision were 6.0 and 2.3% for DMPD, respectively, and -10.4 and 17.1%, for DBS, respectively. The DMPD technique performed better than regular DBS by eliminating the hematocrit effect related blood volume bias. Although this effect was not observed with DMPD, a systematic error of 6.0% was detected and further technical development of DMPD could improve the performance.