Abstract

BackgroundVancomycin is a widely used antibiotic for the treatment of gram-positive bacterial infections, especially for methicillin-resistant Staphylococcus aureus (MRSA) infections. Due to a small therapeutic range and large inter-patient variability, therapeutic drug monitoring (TDM) of vancomycin is required to minimize toxicity and maximize treatment efficacy. Venous blood sampling is mostly applied for TDM of vancomycin, although this widely used sampling method is more invasive compared to less painful alternatives, such as the dried blood spot (DBS) method, which can be performed at home. MethodWe developed an UPLC-MS/MS method for the quantification of vancomycin and creatinine in DBS. A fast sample preparation and short analysis run time of 5.2 min were applied, which makes this method highly suitable for clinical settings. Validation was performed according to international (FDA and EMA) guidelines. ResultsThe validated concentration range was found linear for creatinine from 41.8 µmol/L to 722 µmol/L and for vancomycin from 3.8 mg/L to 76.6 mg/L (r2 > 0.990) and the inaccuracies, imprecisions, hematocrit effects, and recoveries were < 15 % for both compounds. No significant carryover effect was observed. ConclusionHence, we successfully validated a quantification method for the simultaneous determination of creatinine and vancomycin in DBS.