Abstract

To compare the drainage efficiency of double-lumen sump catheters and single-lumen catheters in an in vitro model. Four 12-F, 27.5-cm catheters were used in the study. Three of these had a double lumen with a 7-F internal diameter drainage lumen and one had a single lumen with an 8-F internal diameter drainage lumen. The sump hole opened internally to the drainage lumen in one catheter, opened externally in one catheter, and was occluded in one catheter. Catheters were connected to continuous low-wall suction and were immersed in water, viscous fluid, and viscous fluid with particles. The volume of fluid drained by each catheter was measured and compared. The single-lumen catheter with an 8-F internal diameter drainage lumen drained a volume equal to or larger than that for double-lumen catheters with a 7-F internal diameter drainage lumen for each type of fluid. The open sump of the double-lumen catheters may have decreased the catheter's drainage efficiency by allowing air to be suctioned into the catheter. Single-lumen catheters performed as well as or more efficiently than double-lumen catheters of the same outer diameter in this in vitro model.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.