Abstract
The spatial resolution in far-field mid-infrared (λ>2.5 µm) microscopy and micro-spectroscopy remains limited with the full-width at half maximum of the point-spread function ca. λ/1.3; a value that is very poor in comparison to that commonly accessible with visible and near-infrared optics. Hereafter, it is demonstrated however that polymer beads that are centre-to-centre spaced by λ/2.6 can be resolved in the mid-infrared. The more than 2-fold improvement in resolution in the far-field is achieved by exploiting a newly constructed scanning microscope built around a mid-infrared optical parametric oscillator and a central solid-immersion lens, and by enforcing the linear polarization unidirectional resolution enhancement with a novel and robust specimen error minimization based on a particle swarm optimization. The method is demonstrated with specimens immersed in air and in water, and its robustness shown by the analysis of dense and complex self-assembled bead islands.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.