Abstract

The problem of correct eye aberrations measurement is very important with the rising widespread of a surgical procedure for reducing refractive error in the eye, so called, LASIK (laser-assisted in situ keratomileusis). The double-pass technique commonly used for measuring aberrations of a human eye involves some uncertainties. One of them is loosing the information about odd human eye aberrations. We report about investigations of the applicability limit of the double-pass measurements depending upon the aberrations status introduced by human eye and actual size of the entrance pupil. We evaluate the double-pass effects for various aberrations and different pupil diameters. It is shown that for small pupils the double-pass effects are negligible. The testing and alignment of aberrometer was performed using the schematic eye, developed in our lab. We also introduced a model of human eye based on bimorph flexible mirror. We perform calculations to demonstrate that our schematic eye is capable of reproducing spatial-temporal statistics of aberrations of living eye with normal vision or even myopic or hypermetropic or with high aberrations ones.