Abstract

We propose an aberration-free monochromatic x ray backlit imaging scheme using a combination of convex and concave spherically bent crystals. This configuration works with a wide range of Bragg angles, satisfying the conditions for stigmatic imaging at a particular wavelength. However, the assembly accuracy of the crystals must meet the Bragg relation criteria for spatial resolution to increase the detection efficiency. Here, we develop a collimator prism with a cross reference line engraved on a plane mirror to adjust a matched pair of Bragg angles as well as the intervals between the two crystals and the object to be coupled with the detector. We explore the realization of monochromatic backlighting imaging with a concave Si-533 crystal and a convex α-Quartz-2023 crystal, obtaining a spatial resolution of approximately 7 µm and a field of view of at least 200 µm. To the best of our knowledge, this is the best spatial resolution of monochromatic images of a double-spherically bent crystal to date. Our experimental results are presented to demonstrate the feasibility of this imaging scheme with x rays.