Abstract

We simulate and demonstrate experimentally an inner-wall grating double slot micro ring resonator (IG-DSMRR) with a center slot ring radius of only 6.72 µm based on the silicon-on-insulator platform. This novel photonic-integrated sensor for optical label-free biochemical analysis boosts the measured refractive index (RI) sensitivity in glucose solutions to 563 nm/RIU with the limit of detection value being 3.7 × 10-6 RIU (refractive index units). The concentration sensitivity for sodium chloride solutions can reach 981 pm/%, with a minimum concentration detection limit of 0.02%. Using the combination of DSMRR and IG, the detection range is enlarged significantly to 72.62 nm, three times the free spectral range of conventional slot micro ring resonators. The measured Q-factor is 1.6 × 104, and the straight strip and double slot waveguide transmission losses are 0.9 dB/cm and 20.2 dB/cm, respectively. This IG-DSMRR combines the advantages of a micro ring resonator, slot waveguide, and angular grating and is highly desirable for biochemical sensing in liquids and gases offering an ultra-high sensitivity and ultra-large measurement range. This is the first report of a fabricated and measured double-slot micro ring resonator with an inner sidewall grating structure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.